The A16 AED from Mediana is available in both fully and semi-automatic options. With 3 pre-loaded languages, English, Mandarin and Korean, the ability to change from adult to child mode at the touch of a button as well as being rated to IP55, it is ideal for use in a variety of locations and environments such as workplaces, schools or for public use. The unit has also been drop tested to 0.75m on all 6 sides, proving its robustness.
Key Functions and Features
- PACE-PULSE DETECTION AND REJECTION FUNCTION: The A16 filters external pulse, like a pace-maker, and removes from the ECG reading.
- FAST SHOCK DELIVERY: Shock delivery within 10 seconds.
- MULTILINGUAL: English, Chinese (Mandarin), Korean.
- OPERATION: Fully-automatic AED – no need to press the shock button. Semi-automatic AED – flashing button indicates ‘ready for shock delivery’ users press the button to deliver shock.
- VOICE RECORDING: Enables a check to see what kind of on-site checks were done. When the device turns on the A16 records automatically.
- PADS QUALITY CHECK: Checks connectivity, expiry date or whether they have already been used and damaged.
AED Specifications
DEFIBRILLATION ELECTRIC SHOCK | |
---|---|
Waveform: | Biphasic Truncated Exponential (BTE) waveform (Impedance compensation) |
Energy: | Adult: 170 to 195J (±5%) Paediatric: 44 to 51J (±5%) |
Operating Mode: | Semi-Auto, Fully-Auto |
ECG | |
Lead: | II (RA, LL) |
Patient Impedance: | 25 to 200 ohm |
Detection V/F: | ≥ 0.2mV |
V/T Adult: | ≥ 150 bpm |
INDICATION | |
CONTROLS | |
Standard: | Power button, shock button, patient mode switch button, select language button. |
INDICATORS | |
Visible: | Action icon, Status LCD (AED, battery, temperature and pads status). |
Audible: | Audio speaker (voice prompt, CPR indication). Beep (CPR indication, power on, critically low battery, self-test fail, alarm of abnormal operation). |
PHYSICAL | |
Dimensions: | 286.5H × 200W × 90mmD |
Weight: | Approx. 1.95kg (Including battery, excluding pads). |
Ingress Protection: | IP55 |
SELF TEST | |
Cycle: | Every 24 hours, 1 week, 1 month. Power on self test, battery insertion self test. |
Test Result: | Status LCD displays ‘O’ / ‘X’. |
DATA BACKUP AND COMMUNICATION | |
Standard: | Internal memory, SD card slot, IR communication port. |
Accessory Specifications
PADS | |
---|---|
ADULT/PAEDIATRIC | |
Shelf Life: | 36 months from manufacture date (ref 2902) |
Electrodes: | Disposable pads |
Placement: | Adult: Anterior–lateral Paediatric: Anterior–posterior |
Minimum active gel area: | 80 cm² ±5% |
Cable Length: | Approx. 1.8 m |
ENVIRONMENTAL CONDITIONS — PADS | |
Temperature: | Operation: 0 to 43 °C (32 to 109.4 °F) Storage: 0 to 43 °C (32 to 109.4 °F) |
Relative Humidity: | 5 to 95% RH (non-condensing) |
BATTERY | |
Type: | LiMnO₂, disposable, long-life primary cell |
Voltage/Capacity: | 15 V, 3000 mAh |
Shelf Life: | 2 years from date of manufacture (in original packaging) |
Standby Life: | 5 years from date of manufacture (inserted in the AED) |
Discharge: | A minimum of 200 shocks (excluding the CPR period between defibrillation therapy) or more than 6 hours of operating time at 20 °C |
ENVIRONMENTAL CONDITIONS — BATTERY | |
Temperature: | Operation: 0 to 43 °C (32 to 109.4 °F) Storage: 10 to 25 °C (50 to 77 °F) |
Relative Humidity: | 5 to 95% RH (non-condensing) |
The pad configuration must always match device configuration. | |
Information correct at time of going to press. Subject to change without notice. |
Q&A's
Q: Can this be used for both adults and children?
A: Yes, and the mode can be changed using the adult to paediatric mode button on the main interface.
Q: What languages are the unit’s voice prompts in?
A: There are 3 languages loaded onto the A16, these are English, Chinese (Mandarin), Korean.
Q: What temperature can the unit be stored at?
A: Between 0° and 43° Celsius.
Q: What IP is this rated to?
A: The A16 unit is rated to IP55.
Q: How many years of warranty does the unit have?
A: The A16 has 5 years of warranty.
Q: What is the difference between the two types of pads?
A: Code 2902 are the normal adult/paediatric pads. Code 2903 includes a “pads quality” function, which checks whether there is a problem with the pads, for example, expired pads, damaged pads or the recommended replacement of the pads.

© 2025 Medstock Pty Ltd

Medstock acknowledges the Traditional Owners of Country throughout Australia. We pay our respects to Elders past, present and future.
Customer Service
About Medstock


Customer Service
About Medstock
Open Monday to Friday 9am to 5pm
1300 652 763

Medstock acknowledges the Traditional Owners of Country throughout Australia. We pay our respects to Elders past, present and future.
© 2025 Medstock Pty Ltd